Elab6600N Total Nitrogen Elemental Analyzer Products Model:Elab6600N Origin: China Price: Negotiable Request for Information Consultation Price |
1. Elab6600N can determine solid samples immediately after determining volatile liquid samples without
changing quartz tube and injector. Using only one sampling device, it can measure the total nitrogen
content in gas, liquid (including volatile liquid) and solid samples without adding or replacing any
device.
2. Elab6600N has excellent repeatability and reproducibility of measurement results for measuring
samples of µg/L concentration.
3. Ultra-wide linear range: 0.005mg/L-5000mg/L;4. Elab6600N adopts modular design, so that the instrument can be used as a two-element
combination analyzer, such as: total sulfur and total nitrogen analyzer (Elab6600SN), etc.
5. Repeatability: The total nitrogen analyzer measures the standard substance for light oil
nitrogen content (Standard Substance Number: GBW(E) 060111;standard value: 0.2mg/L),
with an injection volume of 2μL (The actual injection volume is 2μL). The RSD of no less
than 5 consecutive measurements (data cannot be deleted in the process of measuring) is
≤ 10%.
6. Linear: The total nitrogen analyzer measures the "standard substance for light oil nitrogen
content" ((Standard Substance Number: GBW(E) 060111;standard values: 0.2mg/L, 0.5mg/L,
1.0mg/L), with an injection volume of 2μL (The actual injection volume is 2μL). Each
standard sample is measured continuously for 5 times (data cannot be deleted in the process
of measuring). The average of the 5 integrated values of the three standard samples (0.2mg/L,
0.5mg/L, 1.0mg/L) is fitted once, and the linear correlation coefficient is ≥ 0.999. The integrated
value of the 0.5mg/L nitrogen standard sample is at least twice that of the 0.2mg/L nitrogen
standard sample.
Purpose: Used for measuring the total nitrogen content in gas samples, such as propylene gas,liquefied gas, etc; as well as liquid samples such as naphtha, gasoline, kerosene, diesel, benzene, toluene, etc.
Compliance with standards and applicable methods: The total nitrogen analyzer strictly follows the ASTM D4629-17 and SH/T 0657-2007 standards.
Element Determined |
Nitrogen |
Detector |
Chemiluminescence Detector |
Measuring Range |
0.005mg/L—5000mg/L |
Gas Supply |
Oxygen:≥99.995%; Argon:≥99.996% |
Sampling Volume |
Liquid:1—80μl;Gas:1-00ml;Solid:1-10mg |
Electricity Supply |
220VAC±10V,50Hz±0.5 Hz,1000W |