Elab6600S Total Sulfur Elemental Analyzer Products Model:Elab6600S Origin: China Price: Negotiable Request for Information Consultation Price |
1. Elab6600S can determine solid samples immediately after determining volatile liquid samples without
changing quartz tube and injector. Using only one sampling device, it can measure the total nitrogen
content in gas, liquid (including volatile liquid) and solid samples without adding or replacing any
device.
2. Elab6600S has excellent repeatability and reproducibility of measurement results for measuring
samples of µg/L concentration.
3. Ultra-wide linear range: 0.005mg/L-5000mg/L;4. Elab6600S adopts modular design, so that the instrument can be used as a two-element
combination analyzer, such as: total sulfur and total nitrogen analyzer (Elab5600SN), etc.
5. Repeatability: The total sulfur analyzer determines the standard substance for light oil
sulfur content (Standard Substance Number: GBW(E) 060108;standard value: 0.2mg/L),
with an injection volume of 2μL (The actual injection volume is 2μL). The RSD of no less
than 5 consecutive measurements (data cannot be deleted in the process of measuring)
is ≤ 10%.
6. Linear: The total sulfur analyzer measures the "standard substance for light oil sulfur
content" (Standard Substance Number: GBW(E) 060108;standard values: 0.2mg/L,
0.5mg/L, 1.0mg/L), with an injection volume of 2μL (The actual injection volume is
2μL). Each standard sample is measured continuously for 5 times (data cannot be
deleted in the process of measuring). The average of the 5 integrated values of the
three standard samples (0.2mg/L, 0.5mg/L, 1.0mg/L) is fitted once, and the linear
correlation coefficient is ≥ 0.999. The integrated valueof the 0.5mg/L sulfur
standard sample is at least twice that of the 0.2mg/L sulfur standard sample.
Purpose: Used for measuring the total sulfur content in gas samples, such as propylene gas, liquefied gas, etc; as well as liquid samples such as naphtha, gasoline, kerosene, diesel, benzene, toluene, etc.
Compliance with standards and applicable methods: The total sulfur analyzer strictly follows the standards of ASTM D5453-19a, ASTM D6667-21, SH/T 0689-2000, and GB/T 11141-2014.
Element Determined |
Sulfur |
Detector |
UV Fluorescence Detector |
Measuring Range |
0.005mg/L—5000mg/L |
Gas Supply |
Oxygen:≥99.995%; Argon:≥99.996% |
Sampling Volume |
Liquid:1—80μl;Gas:1-00ml;Solid:1-10mg |
Electricity Supply |
220VAC±10V,50Hz±0.5 Hz,1000W |