Elab-TOC/WT Total Organic Carbon Analyzer(UV Oxidation Method)
Products Model:Elab-TOC/WT
Origin: China
Price: Negotiable

 Request for Information   Consultation Price 
Introduction:
   Elab-TOC/WT Total Organic Carbon Analyzer is developed and manufactured by Suzhou Elab Analytical Instruments Co., Ltd. Independently. It uses UV Oxidation non-dispersive Infrared detection method to determine the contents of total organic carbon, and it accords with Chinese Pharmacopoeia of 2010 edition.
Outstanding Features:
1.Elab-TOC/WT has the unique electronic refrigeration module and membrane dehydration technology,
   and its excellent dehydration technology ensures the dehydrated efficiency of the whole system.
   This technology has obtained the national patent!

2.Fast Determination——less than 1.5 minutes for determining TIC, less than 2.5 minutes for determining
   TC.

3.The heated IC inorganic carbon reaction tank eliminates sample peak tailing, and this technology has
   obtained the national patent!

4.Elab-TOC/WT uses total reflection ultraviolet strong oxidation reaction tank so that it has excellent
   system adaptability. It can make the sample transformed completely without adding any other
   oxidizing reagent, avoiding the blank interference caused by adding oxidizing reagents during the
   measurement process. This technology has obtained the national invention patent!

Technical Parameter

Detection Method

UV Oxidation- Non-Dispersive Infra-Red (NDIR) Method

Detector

NDIR

Measurement Range of TOC

0—1000mg/kg

RSD of TOC

≤1.5%

Dehydration

Electronic Refrigeration Module Dehydration & Membrane Dewatering Device

Catalyst

Catalyst Unwanted

Gas

Argon/Nitrogen:≥99.995%

Sampling Volume

Liquid:1—10ml

Power Supply

220VAC±10V,50Hz±0.5 Hz,500W