中文|English
Products
 
Elab-TOC/DT Total Organic Carbon Analyzer(Combustion Oxidation Method)
Products Model:Elab-TOC/DT
Origin: China
Price: Negotiable

 Request for Information   Consultation Price 
Introduction:
   Elab-TOC/DT Total Organic Carbon Analyzer is developed and manufactured independently by Suzhou Elab Analytical Instruments Co., Ltd. It uses combustion oxidation non-dispersive Infrared detection method to determine the contents of total organic carbon, and it accords with International Standard: ISO8245, National Environmental Protection Standard of China: HJ501-2009, National Metrology Verification Regulation of China: JJG821-2005, as well as Chinese Pharmacopoeia of 2010 edition. The emergence of Elab-TOC/DT analyzer filled the blank area of domestic TOC analyzer using high combustion oxidation Non-dispersive Infrared Detection method to determine total organic carbon.
Outstanding Features:
1.Elab-TOC has the unique electronic dual cooling module and membrane dehydration technology. The
   excellent three-stage dehydration technology of the instrument ensures the dehydrated efficiency of
   the whole system,and this technology has been patented!

2.Fast Determination——less than 1.5 minutes for determining TIC, less than 1.5 minutes for determining
   TC.

3.The heated IC inorganic carbon reaction tank eliminates sample peak tailing, and this technology has
   been patented!

4.Elab-TOC can measure liquid (water), solid (junk and soil) samples without replacing any hardware by
   using the same pyrolysis combustion system. The special designed quartz pyrolysis tube and the liquid
   -solid sampler have obtained national invention patents!

5.The unique gas circuit buffer design ensures the excellent reproducibility of the sample measurement
   results,the infrared sensor monitoring device ensures the safety of the system,and these
   technologies have been patented!

6. Equipped with a liquid automatic sampler, the automatic detection of samples can be realized.
Technical Parameter

Detection Method

Combustion Oxidation- Non-Dispersive Infra-Red (NDIR) Method

Detector

NDIR Detector

Measurement Range of TOC

0-1000mg/kg

RSD of TOC

≤1.5%

Highest Salt-Tolerance

100g/L

Dehydration

Three-stage Dehydration—Electronic Dual Cooling Module Dehydration & Membrane Dehydration Unit

Catalyst

Pt/CeO

Gas

Oxygen:≥99.995%

Sampling Volume (μl)

Liquid:1-1000μl

Power Supply

220VAC±10V,50Hz±0.5 Hz,1000W